The hot plate nociception test method is a commonly used method in analgesic screening and detection, and it is also a method that can determine and distinguish the analgesic mechanisms of central nerves and peripheral nerves, with a wide range of applications. In the past, due to imprecise temperature control, inaccurate timing and operational errors by personnel during use, the results of experiments were often of low specificity. In response to the above problems, we have developed the rat and mouse hot plate nociception tester to solve the existing problems in the hot plate method, which has been highly praised by experts and professors during trial and practical use.
Technical Parameters
1. Time to heat from room temperature to 55℃: ≤5 minutes
2. Temperature display accuracy: 0.1℃
3. Temperature control accuracy: 0.1℃
4. Time display accuracy: 0.01s
5. Time display range: 0.01 seconds ~ 99 minutes 59.99 seconds
6. Temperature sensor: Digital temperature sensor
7. Temperature sensor resolution: 0.0625℃
8. Temperature setting range: Room temperature ~ 80℃
9. Hot plate diameter: 170mm
10. Display mode: 5-inch touch screen, all indicators displayed simultaneously
11. Display content: Actual temperature, set temperature, pain threshold time, output, etc.
12. Data output: With USB interface, data can be stored externally
13. Internal data storage capacity: 500 groups of data
14. Recording method: Foot switch, touch switch
15. Operating temperature: 0℃ ~ 45℃
16. Relative air humidity: 15 ~ 80%RH
17. Power supply voltage: AC110~220V±20V
18. Power supply frequency: 50Hz±2Hz
19. Input power: ≤300W
20. Main unit dimensions: 260*260*155mm
21. Weight: 3Kg
简体中文
English